ILAC G8 Decision Rule Visualiser

ILAC G8 · Decision Rule Visualiser

Simple acceptance & guard-band rules — normalised measurement axis

How to use this tool:


1. Select decision rule.

2. Adjust uncertainty and tolerance values

3. Use the measured value (Ym) slider to move the value

4. Read the instrument status statement (Pass, Fail, Pass*, Fail*)

5. Read the associated percentages for tolerance status

Ym 3.20
±T 4.00
um 0.800
Um (k=2) 1.600
TUR (T / Um) 2.50
Guard band w 0.000
Accept limit A = T − w 4.000
✓ PASS
Pass: |Y| ≤ A
Pass*: A < |Y| ≤ T
Fail*: T < |Y|, |Y|−w ≤ T
Fail: |Y|−w > T
PDF g(η)
Pass zone (≤A)
Pass* zone (A–T)
Fail* zone (T–T+w)
Fail zone (>T+w)
Statement: Instrument In Tolerance
Correct

False Accept (PFA)
Statement: Instrument Out Of Tolerance
False Reject (PFR)

Correct
-
ILAC G8:09/2019 · EA-4/02 M:2022 · Harper Metrology